Liaoning Huake Petroleum Apparatus Science & Technology Co., Ltd.

> Products > Others > HK-18854-L2 Automatic Oil Particle Counter (Benchtop Type)

HK-18854-L2 Automatic Oil Particle Counter (Benchtop Type)

Brand : HK
Product origin : China
Delivery time : within 30 days after recieving full TT payment
Supply capacity : 100set/month

The HK-18854-L2 Automatic Oil Particle Counter (Benchtop Type) complies with the requirements of standards such as GB/T 18854 and ISO 11171. It is developed based on the light blockage (light shading) method counting principle and fully conforms to military standards, national standards, and international standards. It can provide fast, accurate, reliable, and repeatable test results as well as complete contamination monitoring and analysis reports. It is suitable for testing the particle contamination level of hydraulic oil, lubricating oil, fire-resistant oil, insulating oil, turbine oil, and other oils. It can be widely used in fields such as aviation, aerospace, electric power, petroleum, chemical industry, transportation, ports, metallurgy, machinery, and automobile manufacturing.

Overview

The HK-18854-L2 Automatic Oil Particle Counter (Benchtop Type) complies with the requirements of standards such as GB/T 18854 and ISO 11171. It is developed based on the light blockage (light shading) method counting principle and fully conforms to military standards, national standards, and international standards. It can provide fast, accurate, reliable, and repeatable test results as well as complete contamination monitoring and analysis reports. It is suitable for testing the particle contamination level of hydraulic oil, lubricating oil, fire-resistant oil, insulating oil, turbine oil, and other oils. It can be widely used in fields such as aviation, aerospace, electric power, petroleum, chemical industry, transportation, ports, metallurgy, machinery, and automobile manufacturing.

Functional Features

  1. 1.Adopts the light blockage (light shading) method counting principle formulated by the International Committee for Hydraulic Standards.

  2. 2.Equipped with a high-precision laser sensor, featuring a wide test range, stable performance, low noise, and high resolution.

  3. 3.Uses a high-pressure injection pump sampling method, allowing self-setting of sampling volume, with stable sample injection speed and high sampling accuracy.

  4. 4.The sample injection system combining positive and negative pressure can realize sample degassing, suitable for testing samples with different viscosities.

  5. 5.Built-in pressure sensor, enabling direct setting of pressure value and automatic judgment of air pressure in the chamber to ensure safety.

  6. 3.The main unit has a built-in air purification system to prevent test contamination, with high integration level, avoiding secondary pollution caused by poor air tightness between the air pump and the test system.

  7. 7.Built-in multiple calibration curves, compatible with all common domestic and foreign standards for calibration.

  8. 8.Built-in more than ten common standards including GJB 420A, GJB 420B, NAS 1638, GB/T 14039, ISO 4406, SAE 4059cpc, SAE 4059F, SAE 749D, ГOCT 17216, QC/T 29104, JB/T 9737, DL/T 432, and HH 005-2018.

  9. 9.Supports custom standard testing, allowing naming of custom standards and grade determination.

  10. 10.Up to 64 detection channels can be set simultaneously according to customer needs.

  11. 11.Built-in data analysis system, which can provide test data and contamination level of all standards with one test.

  12. 12.Capable of cleaning detection; preset the target particle concentration for cleaning requirements, display the cleaning status in real-time during the cleaning process, and automatically stop when the requirements are met.

  13. 13.Built-in viscosity, moisture, and temperature sensor modules (optional), which can provide reference values of viscosity, moisture content saturation, ppm value, and temperature while accurately testing particle distribution.

  14. 14.Particle size can be set arbitrarily, with nearly 10,000 built-in particle sizes for easy particle size analysis.

  15. 15.Multiple sampling containers such as standard sampling bottles or sampling cups can be used to meet the testing requirements of different industries.

  16. 16.Equipped with a full-function 7-inch color touch screen for operation, with a foldable button design, supporting Chinese and English dual interfaces, and easy to operate.

  17. 17.Capable of manual, automatic, and semi-count calibration.

  18. 18.Equipped with an RS232 interface, supporting 485 communication, and can be connected to a computer or laboratory platform for data processing.

  19. 19.Has massive data storage and printing functions, supports USB flash drive storage, and the data stored in the unit can be exported to a USB flash drive.

  20. 20.Allows setting of more than 5 accounts, and separate setting of operator permissions.

Technical Parameters

  1. Light Source: Semiconductor laser

  2. Particle Size Range: 0.8μm~600μm (calibrated as 1μm~100μm according to ACFTD or 4μm~70μm(c) according to ISO MTD)

  3. Detection Channels: 8~64 channels optional (64 channels by default at delivery), with arbitrary particle size setting

  4. Sensitivity: 0.8μm (ISO4402) or 3μm (c) (GB/T18854, ISO11171)

  5. Resolution: Better than 6%

  6. Repeatability: RSD<2%

  7. Sample Test Viscosity: ≤650cSt (for samples with high viscosity, heating or dilution methods can be used for testing)

  8. Sampling Volume: 0.2mL~6000mL, with an interval of 0.1mL

  9. Sampling Accuracy: Better than ±0.2%

  10. Sampling Speed: 5mL/min~80mL/min

  11. Air Pressure Chamber: Positive and negative pressure chamber device, realizing sample degassing and testing of high-viscosity samples

  12. Maximum Vacuum of Air Pressure Chamber: 0.08MPa

  13. Maximum Positive Pressure of Air Pressure Chamber: 1MPa

  14. Maximum Particle Concentration: 12000 particles/mL~40000 particles/mL

  15. Temperature (Paid Optional): Collection range: 1℃~100℃; Collection accuracy: 1℃

  16. Water Activity (Paid Optional): Collection range: 1%RH~100%RH; Collection accuracy: 1%RH

  17. Moisture Content (Paid Optional): Collection range: 1ppm~360ppm; Collection accuracy: 1ppm

  18. Test Sample Temperature: 0℃~80℃

  19. Operating Temperature: -20℃~60℃

  20. Storage Temperature: -30℃~80℃

  21. Power Supply: AC 110V~245V 50/60Hz, 70W

  22. Main Unit Dimensions: (340×410×650)mm

  23. Air Pump Dimensions: (180×160×220)mm

  24. Net Weight of Main Unit: ≯23kg; Net Weight of Air Pump: ≯5.5kg


Please feel free to give your inquiry in the form below. We will reply you in 24 hours.
Message