Overview
Functional Features
1.The oil cup adopts a three-electrode structure with an inter-electrode distance of 2mm, which can eliminate the influence of stray capacitance and leakage on dielectric loss test results.
2.The instrument uses medium-frequency induction heating and a PID temperature control algorithm to strictly control the temperature within the preset temperature error range.
3.The internal standard capacitor is an SF6-gas-filled three-electrode capacitor, ensuring the instrument's accuracy is maintained even after long-term use.
4.The AC test power supply adopts an AC-DC-AC conversion method, effectively avoiding the impact of mains voltage and frequency fluctuations on the accuracy of dielectric loss testing.
5.Comprehensive protection functions: when overvoltage, overcurrent, or high-voltage short circuit occurs, the high voltage is quickly cut off and a warning is issued; when the temperature sensor fails or is not connected, a warning message is sent.
6.Convenient test parameter setting: temperature setting range is 0~120℃, AC voltage setting range is 500V~2200V, and DC voltage setting range is 300~500V.
7.Equipped with a large LCD screen with backlight for clear display; it has a user-friendly interface and can automatically store and print test results.
8.Built-in real-time clock; the test date and time can be saved, displayed, and printed together with the test results.
9.Empty electrode cup calibration function; calibration data is automatically saved to facilitate the accurate calculation of relative permittivity and DC resistivity.
Technical Parameters
Power Supply: AC 220V±10%, 50Hz/60Hz±1%
Power Consumption: 100W
Measurement Range:
1.Capacitance: 5pF~200pF
2.Relative Permittivity: 1.000~30.000
3.Dielectric Loss Factor: 0.00001~100
4.DC Resistivity: 2.5MΩ·m~20TΩ·m
Measurement Accuracy:
1.Capacitance: ±(1% reading + 0.5pF)
2.Relative Permittivity: ±1% reading
3.Dielectric Loss Factor: ±(1% reading + 0.0001)
4.DC Resistivity: ±10% reading
Resolution:
1.Capacitance: 0.01pF
2.Relative Permittivity: 0.001
3.Dielectric Loss Factor: 0.00001
Temperature Measurement Range: 0~120℃
Temperature Measurement Error: ±0.5℃
AC Test Voltage: 500V~2200V, continuously adjustable, frequency 50Hz
DC Test Voltage: 300V~500V, continuously adjustable
Environmental Requirements: Temperature 0℃~40℃; Humidity <75%
Dimensions and Weight: (500×360×420)mm, 22kg